Specifications
· Infusion Rate : 0.1 ml/h—1200m1/h (1m1/h step)
· Rate Accuracy : within ±5% (after correct calibration)
· Purge Rate : 100m1/h—1200m1/h (can be calibrated)
· Volume Limit : 0ml—9999m1
· Total Infusion Volume : 0—9999.9m1
· Alarm Pressure : High : 800mmHg±200mmHg (106.7kPa±26.7kPa)
Medium : 500mmHg±100mmtlg (66.7kPa±13.3kPa)
Low : 300mmHg±200mmHg (40.7kPa±13.3kPa)
· Alarm : End of injection, occlusion, door open, bubbles in the tube,
wrong setting, low battery, AC power pulled out etc.
· Power Source : 100V—240V 50/60Hz;4 hours internal battery backup
· KVO Speed : lml/h—5m1/h (1m1/11 step)
· Power Consumption : 18VA
· Operation Environment : Temperature +5°—-+40°
Relative Humidity 20—-90%